EasyManuals Logo

Bruker M4 TORNADO User Manual

Bruker M4 TORNADO
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #98 background imageLoading...
Page #98 background image
Using the M4 TORNADO Software
98
User Manual
6.8 Line Scan Analysis
The Line workspace allows the element distribution analysis along a line.
Fig. 40 The Line workspace
6.8.1 Definition of measurement conditions
The measurement conditions can be set on the right side of the dialog. The display of the sample
image is similar to Point analysis (see section 6.5.1).
Use the mouse to drag both ends of the scan line to the desired positions. Information on the
image scan and spectrum windows can be found in the corresponding sections.
Table 34 Functions of the Line workspace
Object Description Hints
Display of length of the line scan
Set number of points along the scan
line or the distance between points.
The minimum step size is 4 µm.
Both selections will alter. The
value for the other selection is
calculated by the chosen value..
Display of the distance between
points and X-ray spot diameter
The distance between points
will change depending on
whether Points or Distance
has been selected.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker M4 TORNADO and is the answer not in the manual?

Bruker M4 TORNADO Specifications

General IconGeneral
BrandBruker
ModelM4 TORNADO
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals