Visual-ICE™ Cryoablation System
7-23
Message Reason for Appearance
60-03 Network Connection
• You are connected to the Internet
• However, there is no connection with the server at
Galil Medical
• If problem persists, contact Galil Medical Customer
Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
The Visual-ICE System found the
Internet but not Galil Medical servers.
The user needs to call Galil Medical
because the problem is on the Galil
Medical side.
60-04 Update Failed
• An error occurred during the update process
• Retry update
During User update of the system, an
error occurred that prevented completion
of the update. Another update should be
attempted.
60-05 Incompatible Hardware
• The hardware is not compatible with the current
software
• Try downloading a software update from the Galil
Medical servers or from a Galil Medical USB ash
drive
When the system performed self-
checks at start up, an incompatibility
between the hardware and software was
detected.