7-24 User Manual
Message Reason for Appearance
60-06 Incompatible Software
• The Visual-ICE System software is not compatible
with regulatory approvals
• Please connect to the Galil Medical network to
download the appropriate software for your market
• If problem persists, contact Galil Medical Customer
Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
The software was checked against
approved software versions in the
regulatory les for each respective
market. Incompatibility with regulatory
approval was identied. The appropriate
software should be downloaded.
60-07 Invalid Conguration
• A software le is nonoperational
• If problem persists, contact Galil Medical Customer
Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
There was a problem with the software’s
conguration les. Contact Galil Medical.
60-08 DHCP Request
• Error renewing IP address
• Connect Ethernet cable or check the Ethernet port
• Contact your network administrator if unable to
renew
Renew of IP address failed in the
network conguration screen. User
should check the Ethernet connection or
contact their network administrator.