Visual-ICE™ Cryoablation System
7-27
Message Reason for Appearance
SYSTEM
80-01 Communication Failure
• Internal communication failed
• Reconnection attempt failed
• Restarting Visual-ICE System
• If problem persists, contact Galil Medical Customer
Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Software was unable to connect with
hardware after attempt to reinitialize
communication. If the restart fails, the
system is not usable.
80-02 Startup Failure
• System self-checks failed
• Restart the system
• If problem persists, contact Galil Medical Customer
Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Software self-checks located a failure
that required the system to be restarted.
80-03 Pressure Alert
• Pressure exceeds safe operating limits
• Close the gas cylinders
• Procedure will be terminated and gas vented from
the system
System detected the internal pressure
was above safe limits. The system will
terminate the procedure and vent gas
from the system.