Visual-ICE™ Cryoablation System
7-27
Message Reason for Appearance
SYSTEM
80-01 Communication Failure
Internal communication failed
Reconnection attempt failed
Restarting Visual-ICE System
If problem persists, contact Galil Medical Customer
Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Software was unable to connect with
hardware after attempt to reinitialize
communication. If the restart fails, the
system is not usable.
80-02 Startup Failure
System self-checks failed
Restart the system
If problem persists, contact Galil Medical Customer
Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Software self-checks located a failure
that required the system to be restarted.
80-03 Pressure Alert
Pressure exceeds safe operating limits
Close the gas cylinders
Procedure will be terminated and gas vented from
the system
System detected the internal pressure
was above safe limits. The system will
terminate the procedure and vent gas
from the system.