7-28 User Manual
Message Reason for Appearance
80-04 Temperature Warning
• The internal temperature of the Visual-ICE System
exceeds appropriate operating limits
• Discontinue the cryoablation procedure as soon as
safe to do so
• Contact Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
The internal temperature of the system
exceeded appropriate operating limits.
80-05 Service Due
• Low battery detected
• Contact Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
System detected a low battery. System
operation may be impacted if operated
with a low battery.
System Error messages are displayed in
the right corner of the navigation tool bar.
The User is oered details of the error by
“Press here” instruction.
80-30 System Error
• Gas pressure display may be inaccurate. Monitor
procedure carefully using image guidance. At the
conclusion of the procedure, contact Galil Medical
Customer Service.
• Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Internal pressure checks were
inconsistent and may produce an
inaccurate display on the Pressure
Gauge.