Message Reason for Appearance
80-31 System Error
• Gas cylinder shut o valve is not open enough to
provide adequate ow. If necessary, open the valve
approximately another half turn.
• Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Gas ow from gas cylinder was
insucient. To improve gas ow, the
cylinder valve should be opened further.
80-32 System Error
• Channel X is defective. Choose another channel.
At the conclusion of the procedure, contact Galil
Medical Customer Service.
• Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
A defective solenoid on Channel X was
detected; another channel should be
selected.
80-33 System Error
• Defective MTS in channel [X]. Replace with a new
MTS.
• Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
The MTS temperature display was out
of the expected range when initially
connected to the needle connection
panel.