EasyManua.ls Logo

BTG Visual-ICE - Page 112

BTG Visual-ICE
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-30 User Manual
Message Reason for Appearance
80-34 System Error
Fan X failure. At the conclusion of the procedure,
contact Galil Medical Customer Service.
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
Fan X was identied as not operating.
80-35 System Error
The gas ow rate in channel XX exceeds the
system operation requirements and may impact
performance. Reduce the number of active needles.
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
The calculated ow rate for a given
channel exceeded the optimum system
requirement. The number of active
needles should be reduced.
80-36 System Error
The gas ow rate exceeds the system operation
requirements and may impact performance. Reduce
the number of active needles
Galil Medical Customer Service
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
The cumulative ow rate for all channels
exceeded the optimum system
requirement. The number of active
needles should be reduced.

Table of Contents