5-2 User Manual
Setting Description
Cylinder Volume
Select the gas cylinder volume and units of measure according
to the standard in the geographical region. The gas cylinder
volume and units can only be changed by administrative or
service personnel.
Inactivity Timeout
Select the desired duration from 30 to 180 minutes that the
system can be inactive before requiring you to re-enter your
password. The default inactivity timeout is two hours.
Language Select the language in which the software is displayed.
Maximize by
Default
Select a section of the Procedure Screen to be maximized by
default at login.
Low Cylinder Alert
Select the desired reminder interval (0 to 15 minutes) for the
Gas Indicator to display an alert that the estimated remaining
volume in the gas cylinder is low.
Link all channels
Check this box to automatically link all adjacent channels for
simultaneous operation (e.g., 1 and 2, 3 and 4, 5 and 6, …)
Passive thaw
timer count up
Check this box to automatically display elapsed time during
passive thaw. The digital timer reads Stopped and the time
elapsed in passive thaw is indicated.
Active Flush
Unless disabled by removing the check from this box, an
automated active 30 second ush will follow a freeze cycle
conducted at 50% or higher intensity and lasting greater than 3
minutes.
Display Sensor
Temperatures
Check this box for a Channel Status display of the internal
gas temperature at the needle tip during freeze phase and the
estimated temperature range of the needle shaft during active
thaw phases for i-Thaw and CX-type needles.
FastThaw in
Channel Controls
Check this box for display of the FastThaw™ icon in the
Channel Controls when the system is in i-Thaw™ mode and a
FastThaw capable needle is connected.