Visual-ICE™ Cryoablation System
7-13
Message Reason for Appearance
10-06 Incorrect Response
You have not entered the correct response
Contact Galil Medical Customer Service for a
Response to the on screen Challenge
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
User tried to access the Emergency
Login and did not respond correctly
to the challenge. User must contact
Customer Service in order to receive an
emergency login response. This action
does not reset a password.
10-07 Incorrect Response
You have not entered the correct response
Contact Galil Medical Customer Service for a
Response to the on screen Challenge
o +1 877 639 2796 or
o +1 651 287 5000
o support@galilmedical.com
User tried to reset their password
and did not respond correctly to the
challenge. User must contact Customer
Service in order to reset their password.
PROCEDURE
20-01 Cannot Start Test
The gas pressure is too low to begin a procedure
Check that the gas cylinders have enough pressure
to start the procedure
User pressed the Test button when
either gas cylinders were not yet
connected or the gas pressure is below
operating pressure. To proceed, gas
cylinders with su󰀩cient pressure must
be connected.
20-02 Argon Shut O󰀨 Valve
The argon shut o󰀨 valve may be closed
Check and open if necessary
At system start-up, the system detected
gas connected, but no gas was entering
the system. The argon shut o󰀨 valve
may be closed. To proceed, the argon
shut o󰀨 valve must be open.