7-16 User Manual
Message Reason for Appearance
20-12 Automatic Vent
Do you want to automatically vent high pressure gas
from the system?
YES NO
User was provided an option for
automatically venting the high pressure
gas in the system.
20-13 Automatic Vent
Is the gas supply closed?
YES NO
If user selected the option for automatic
gas venting, the user will need to
conrm the gas supply is closed before
automatic venting is activated.
20-14 Automatic Vent
The gas pressure is not dropping
Check that the gas cylinder shut o󰀨 valve is closed
User selected to automatically vent high
pressure gas at the end of a procedure,
but the pressure was not dropping. User
must ensure the shut o󰀨 valve is closed.
20-15 Automatic Vent
Venting is in progress
If needles are still connected, do not unlock channels
or disconnect needles until venting is completed
User selected to automatically vent
the high pressure gas at the end of the
procedure.
20-16 Automatic Vent
Automatic venting successfully completed
User chose to automatically vent the
high pressure gas at the end of the
procedure.