Visual-ICE™ Cryoablation System
7-17
Message Reason for Appearance
20-17 Gas Vent
Before disconnecting the gas hose, manually vent
the Visual-ICE System using the Manual Vent Valve
on the rear of the machine
User selected to not use the feature to
automatically vent high pressure gas
from the System.
20-18 System Shutdown
Do you want to shut down the system?
YES NO
User selected Shutdown in the login
screen to power down the system.
20-19 Procedure Timeout
The procedure has exceeded the allowable time
Procedure will be terminated
The procedure exceeded the allowable 8
hour duration.
20-20 Maximum Activity for FastThaw
Maximum limit for FastThaw reached
Wait until thawing is complete before activating
additional FastThaw needles
This channel will use passive thaw
User attempted to activate more than
four FastThaw needles. The Visual-ICE
System supports simultaneous operation
of no more than four FastThaw needles.
After completing the thaw phase with
four needles, additional needles may be
activated for FastThaw.
20-21 Maximum Activity for i-Thaw
Maximum limit for i-Thaw reached
Wait until thawing is complete before activating
additional i-Thaw needles
This channel will use passive thaw
Thawing with i-Thaw is limited to a
maximum of 9 needles simultaneously
activated in thaw mode.
When thawing with CX-type needles,
thawing is limited to a maximum of 7
needles simultaneously activated.
After completing the thaw phase,
additional needles may be activated..