Visual-ICE™ Cryoablation System
7-17
Message Reason for Appearance
20-17 Gas Vent
• Before disconnecting the gas hose, manually vent
the Visual-ICE System using the Manual Vent Valve
on the rear of the machine
User selected to not use the feature to
automatically vent high pressure gas
from the System.
20-18 System Shutdown
• Do you want to shut down the system?
YES NO
User selected Shutdown in the login
screen to power down the system.
20-19 Procedure Timeout
• The procedure has exceeded the allowable time
• Procedure will be terminated
The procedure exceeded the allowable 8
hour duration.
20-20 Maximum Activity for FastThaw
• Maximum limit for FastThaw reached
• Wait until thawing is complete before activating
additional FastThaw needles
• This channel will use passive thaw
User attempted to activate more than
four FastThaw needles. The Visual-ICE
System supports simultaneous operation
of no more than four FastThaw needles.
After completing the thaw phase with
four needles, additional needles may be
activated for FastThaw.
20-21 Maximum Activity for i-Thaw
• Maximum limit for i-Thaw reached
• Wait until thawing is complete before activating
additional i-Thaw needles
• This channel will use passive thaw
Thawing with i-Thaw is limited to a
maximum of 9 needles simultaneously
activated in thaw mode.
When thawing with CX-type needles,
thawing is limited to a maximum of 7
needles simultaneously activated.
After completing the thaw phase,
additional needles may be activated..