BATTERY SHORTED
122
Battery is not accepting
charge in the initial charge
cycle (the voltage remains
low).
Ensure battery leads are firmly
connected to the battery.
Battery may be fully discharged.
Recharge battery in its original
charger before placing it in the
analyzer, or run Boost, (see
Boost, page 33).
Battery may have shorted cells -
if unable to charge in the original
charger, discard battery.
BATTERY
OVERCHARGED
176
Battery SoC (State of Charge)
is too high for QuickTest or
OhmTest
The battery is discharged for
five minutes, after which
QuickTest restarts. Discharge
is repeated up to 5 times until
SoC is below 90%.
If OhmTest is running, the
program is aborted. Discharge
the battery before running
OhmTest again.
Turn off the automatic OhmTest.
If error continues, battery cannot
be run on QuickTest or OhmTest.
BATTERY
UNDERCHARGED
175
Battery SoC (State of Charge)
is too low for QuickTest or
OhmTest
The battery is charged for five
minutes, after which QuickTest
restarts. Charge is repeated
up to 5 times until SoC is
above 20%.
If OhmTest is running, the
program is aborted. Partially
charge the battery before
running OhmTest again.
BATTERY TOO COLD
12
Battery temperature is too
cold when inserted. The
battery is trickle charged until
temperature rises above 5°C
Allow the battery to warm up.
If the battery is non-OEM, the
thermistor may not match that
on the adapter.