Cadex Electronics Inc. C7000-C Series Battery Analyzers .
124
Detailed
Message
Code Reasons Solutions
CHARGE TIMEOUT
144
Charge time has exceeded the
expected value for the battery.
Verify that the mAh rating of the
battery matches the C-Code mAh
setting.
Analyzer may not have detected
the end of charge conditions.
Reduce the negative slope to
8mV/cell for NiCd and NiMH.
Raise the End of Charge settings
for Li-ion or SLA.
CHARGE WAIT
9
Station is on hold until
sufficient power is available.
This is normal if large batteries
are being serviced. Service
resumes automatically when
sufficient power is available to
the station.
CHARGING
2
Battery is being charged
normally.
RAPID HEAT RISE
154
Midway during charge the
temperature rose too rapidly
(more than 2°C in 2 minutes).
Program stops until battery is
cooled by 5°C and then
resumes. A WAIT code may
appear for the battery to cool
before the next process
begins.
Wait until processing is complete.
Lower the charge rate in the
extended C -Code. Observe
battery performance. The battery
may be old.
CURRENT RISE AT
FULL CHARGE
130
Generally for SLA and Li only.
Battery current suddenly
increased as the end-of-
charge condition was
achieved.
Raise the End of Charge
settings.
If the SLA battery is a Hawker
cell, use the Hawker C-Code
settings (see page 53). Allow
the station to complete service.
CYCLE RESUMED
32
Power resumed or the battery
is re-inserted into the adapter.