Detailed
Message
Code Reasons Solutions
PLATEAU TIMEOUT
113
Battery was fully charged
before full-charge conditions
were met. For SLA and Li
batteries only.
The end-of-charge setting in the
C-Code may have to be
increased. Battery may be old
and operating time less than
manufacturer’s specifications.
POWER ON
200
Power is detected inside the
analyzer.
PROCESS
COMPLETE
15,
35, 36
Current battery-service
program is complete.
PROCESS
SUSPENDED
18
Battery is removed during
service.
Re-insert battery within 5
seconds to resume service. If
message persists, see Code 188
for further information.
SERVICE
INTERRUPTED
188
Battery is removed during
service for more than five
seconds. Program terminated.
If the battery was not removed
the battery protection circuit or
thermistor may have activated
disconnecting the terminals. The
analyzer then thinks that the
battery has been removed.
If the battery is hot, reduce
heating by lowering the Charge
and Discharge. For Li-ion
batteries, reduce the Max Charge
voltage or the Max Standby
Voltage.
PROCESS RESUMING
25
(Code 25) Battery service is
continuing after a power
interruption.
This message appears after the
initial startup screen and
remains until the analyzer
determines the point at which
battery service was interrupted
and resumes the current
program from that point.