Cadex Electronics Inc. C7000-C Series Battery Analyzers .
Detailed
Message
Code Reasons Solutions
CUSTOM PROGRAM
HAS FAILED –
DISPLAY WILL SHOW
REASON FOR
FAILURE
16
(Code 16) The custom
program has produced an
error. The Next statement is
not processed. The program
has been terminated.
See the detailed display for the
exact nature of battery failure.
If Code 16 is displayed, then
there is a bug in the custom
program - verify the custom
program statements.
READY
5
Service has completed and
battery is ready. Final
capacity or state of health is
displayed.
Faults or warnings (if any) were
corrected. Remove battery and
use as normal.
RECONDITION
TIMEOUT
146
Recondition time has
exceeded the expected value
for the battery. Program goes
to the next cycle.
Verify the mAh settings on the C-
Code.
Raise the recondition discharge
rate, (see Recondition Discharge
Rate, page 47).
The battery may be intrinsically
safe, preventing deep discharge.
Use I/S settings for the battery,
(see Intrinsically Safe Batteries,
page 53)
If battery is new or has been in
storage, use the Prime program.
Cells may be mismatched, run
the Prime program.
Monitor the battery use.
RECONDITIONING
4
Battery is being reconditioned.
The program will proceed to the
next step when recondition has
completed.
RESISTANCE TEST
27
A manually selected or
programmed resistance test
for the battery is in progress.
Program completes in
5 seconds.