EasyManua.ls Logo

Cellwatch Frontier - Page 90

Cellwatch Frontier
125 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cellwatch Frontier System Installation & User Manual
Section II – User Guide-90
An example of Frontier monitoring a charge event.
Low float current (continuity)
Frontier includes a low float alarm that is triggered when current measured is less than the intended
float charge. This can be indicative of a battery internal disconnect or possibly switched-off charger. This
limit must be less than the charge current limit, and a Float Charging Current Probe (FCCP) should be
used as the charge CT in a logical CT configuration.
Note: This alarm’s condition is set up specifically for use with FCCP units. This type of current probe
takes 20 minutes for readings to stabilize. Thus, a low float current alarm will not trigger until the
current measured by the FCCP is less than the low float current limit continuously for 20 minutes.
A low float current event takes lower priority to discharge and charge events. This means that a low
float current alarm will never trigger during the event of a discharge or charge. If a low float current
alarm has triggered, the detection of a discharge event or a charge event will cause the low float current
alarm to turn off.

Table of Contents