Manual Operation
NOTICE
1. The table below is the voltage range of OVP.
Model Min. OVP (V) Max. OVP (V)
62000P-xx-xx 1.10 x Vo_SET 1.10 x Vo_MAX
Table 3-4 OVP Range
2. When 0 is set for the OVP, it means the function is disabled.
When OVP occurs the main page will prompt a protection message as shown below:
V _ S E T = 8 0 . 0 0 V I _ S E T = 1 5 . 0 0 A O F F
0 0 . 0 0 V
00 . 00A
0 . 0 W
O V P
Figure 3-58
3.3.5.2 OCP Protection
1. Use
”, “
” keys to move the cursor to the column to be set as Figure 3-59
shows.
[ P R O T E C T I O N ]
O V P = 8 8 . 0 V
O C P = 6 3 . 0 A
O P P = 1 2 6 0 . 0 W
F O L D B A C K = D I S A B L E
Figure 3-59
2. Use the numeric (
0
~
9
) keys or “Rotary” (
) knob to set the value.
3. Press
ENTER
” to confirm.