Programmable DC Power Supply 62000P Series Operating & Programming Manual
[PROGRAM ]
PROG NO . = 1
PROG CHA I N NO. = 0
RUN COUNT = 0
CLEAR SEQ . = NO
[SEQUENCE]
SEQ NO . =1 SEQ . TY P E=AUTO
VOL TAGE= 1 0 ( V ) V S . R . =1 ( V / mS )
CURRENT= 2 0 ( A ) I S . R . =1 ( A / mS )
TTL OUT=1 <BINARY = 00000001>
TIME =5 (S)
[SEQUENCE]
SEQ NO.=2 SEQ. TYPE=MANUAL
VOL TAGE= 8 0 ( V ) V S . R . =1 0 ( V / mS)
CURRENT= 1 5 ( A ) I S . R . =0 . 1 ( A / mS)
TTL OUT=2 <BINARY=00000100>
TIME =10 (S)
[ SEQUENCE ]
SEQ NO . = 9 SEQ. TYPE=AUTO
VOL TAGE= 0 ( V ) V S . R . =1 . 0 ( V / mS )
CURRENT= 0 ( A ) I S . R . =1 . 0 ( A / mS )
TTL OUT= 0 <B INARY=00000000>
TIME =0 .1(S)
[ SEQUENCE ]
SEQ NO . =10 SEQ. TYPE=AUTO
VOL TAGE= 0 ( V ) V S . R . =1 . 0 ( V / mS )
CURRENT= 0 ( A ) I S . R . =1 . 0 ( A / mS )
TTL OUT= 0 <B INARY=00000000>
TIME =0 .1(S)
Figure 4-2
4.1.1 Description of PROGRAM Settings
A PROGGRAM has 4 settings: (1) PROG NO., (2) PROG CHAIN NO., (3) RUN COUNT
and (4) CLEAR SEQ..
4.1.1.1 Setting PROG NO.
1. Use
”, “
” keys to move the cursor to the column to be set as Figure 4-3 (1)
shows.
4-2