Remote Operation
Return Parameter: <NR1>
Query Example: CONF:APGV? It returns the status set.
Return Example: 5V or 10V
5.6.2.3 SOURCE Subsystem
(1) SOURce:VOLTage
Description: It sets the output voltage.
Syntax: SOURce:VOLTage <NRf+>[suffix]
SOURce:VOLTage <NRf+>[suffix]
Parameter: Refer to individual spec for valid numeric range.
Example: SOUR:VOLT 0.01 It sets the output voltage to 0.01 volt.
SOUR:VOLT 80.00 It sets the output voltage to 80.00 volt.
Query Syntax: SOUR:VOLT?
Return Parameter: <NRf+> [Unit Volt]
Query Example: SOUR:VOLT? It returns the voltage.
Return Example: 9.000000e1
(2) SOURce:VOLTage:LIMit:{HIGH/LOW}
Description: It sets the output voltage range.
Syntax: SOURce:VOLTage:LIMIT:HIGH <NRf+>[suffix]
SOURce:VOLTage:LIMIT:LOW <NRf+>[suffix]
Parameter: Refer to individual spec for valid numeric range.
Example: SOUR:VOLT:LIMIT:HIGH 60.0 It sets the output voltage range to
60V maximum.
SOUR:VOLT:LIMIT:LOW 20.0 It sets the output voltage range to
20V minimum.
Query Syntax: SOUR:VOLT:LIMIT:HIGH?
SOUR:VOLT:LIMIT:LOW?
Return Parameter: <NRf+> [Unit Volt]
Query Example: SOUR:VOLT:LIMIT:HIGH? It returns the maximum range set
for voltage.
Return Example: 120.00
(3) SOURce:VOLTage:PROTect:{HIGH}
Description: It sets the voltage range for over voltage protection.
Syntax: SOURce:VOLTage:PROTect:HIGH <NRf+>[suffix]
Parameter: Refer to individual spec for valid numeric range.
Example: SOUR:VOLT:PROT:HIGH 60.0 It sets the high limit to 60V for
voltage output protection.
Query Syntax: SOUR:VOLT:PROT:HIGH?
Return Parameter: <NRf+> [Unit Volt]
Query Example: SOUR:VOLT:PROT:HIGH? It returns the high limit of voltage
protection.
Return Example: 88.00
(4) SOURce:VOLTage:SLEW
Description: It sets the rising or falling slew rate (volt/ms) of output voltage.
5-11