EasyManua.ls Logo

Chroma 62000P Series - Page 88

Chroma 62000P Series
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Operation
NOTICE
1. When entering into the CALILBRATION page, be sure to check the interface connection
on the rear panel is correct and then press “
ENTER
” to start calibration.
2. If Agilent 34401 is used, the DVM1 and DVM2 can be connected to the front and rear
measurement input terminal respectively.
(3) When in the APG Voltage Calibration pages and the connection is correct, press
ENTER
” to confirm.
(4) It will ask users to input about 0.5V voltage signal (Pin 3), and press “
ENTER
” the cursor
will stop at position [1] as Figure 3-98 shows. Adjust the Power Supply to 0.5V±0.2V
and use DVM1 to measure the reading of Power Supply. Input the voltage (2 digits after
decimal point) read by DVM 1 to position [1] and press “
ENTER
” to confirm.
[APG VOLT
A
GE CAL I B
R
A
T I O N ]
C H E C
K
A P G C O N N E C T I O N A N D P
R
ESS [ E N T E
R
]
( S E T ) I N P U T V O L T A G E F O R S E T T I N G = 0 . 5
V
A
C T U
A
L A P G I N P U T V O L T A G E = 0 . 6 6 V
( S E T ) I N P U T V O L T A G E F O R S E T T I N G = 8 . 0
V
A
C T U
A
L A P G I N P U T V O L T A G E = 8 . 6 8 V
]1[
]2[
Figure 3-98
(5) Press
” again will ask users to input about 8.0V voltage signal (Pin 3), and press
ENTER
” the cursor will stop at position [2] as Figure 3-98 shows. Adjust the Power
Supply to 8V±0.2V and use DVM1 to measure the reading of Power Supply. Input the
voltage (2 digits after decimal point) read by DVM 1 to position [2] and press “
ENTER
to confirm.
(6) Press
” again the system will set the output voltage of Pin 5 on the rear panel to
0.5V. The cursor stops at position [3] as Figure 3-99 shows. Input the voltage (2 digits
after decimal point) read by DVM 2 to position [3] and press “
ENTER
” to confirm.
3-59

Table of Contents

Related product manuals