Performing a Test
134
The hourglass icon spins until the sample has
been detected and the test begins. The time
remaining for the test is displayed.
Measurement time:
The time needed to run a test depends on the
test parameter (refer to package insert for
details). Typically, 8 to 12 minutes are
required.
Do not add more blood. Do not touch the test strip until the result is displayed.
For Troponin T only:
As soon as the meter detects a signal (this
may take a few minutes), a qualitative
message (TnT Elevated) appears under the
hourglass icon (see illustration).
The final quantitative result is displayed when
the test is completed. Decisions on further
treatment of the patient should be based on
the final result only. Refer to the Roche
CARDIAC T Quantitative or Roche CARDIAC
POC Troponin T package insert for more
details.
Pat. Test - Measuring
proBNP PB1234
JONES, TOM
Par:
Op:
Code:
SCHULZ, MANFREDPat:
10:20 Min
04/19/2016
Pat. Test - Measuring
Trop T TT 2721
JONES, TOM
Par:
Op:
Code:
SCHULZ, MANFREDPat:
10:20 Min
TnT Elevated
04/19/2016

Table of Contents

Other manuals for cobas h 232

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the cobas h 232 and is the answer not in the manual?

cobas h 232 Specifications

General IconGeneral
Device TypePoint-of-care analyzer
Sample TypeWhole blood, plasma
Measured ParametersTroponin T, Myoglobin, D-Dimer, NT-proBNP
ConnectivityUSB
Intended Usecardiac markers
Data ManagementLIS connectivity
ThroughputSingle test
Sample Volume150 µL
Operating Temperature15-30 °C
Operating Humidity10-85% (non-condensing)
DisplayColor touchscreen
Measurement PrincipleElectrochemiluminescence immunoassay (ECLIA)