Control Testing and Quality Control
148
Always place the meter on a level, vibration-free surface when applying the sample to the
sample application area of the test strip. Once the sample has been completely absorbed by
the test strip, you can move the meter.
CAUTION
10 Using the pipette, draw up the dissolved
contents of the vial.
11 Apply the sample (exactly 150 L) to the
application area of the test strip.
12 Touch to confirm that the sample has
been applied.