Control Testing and Quality Control
147
9 If the control material box contains more
than one level, select the level of choice
for this measurement.
The thermometer icon shows that the test
strip is warming up. A beep tone indicates
when the warming-up process is complete
(provided the beeper is enabled).
The pipette icon indicates that the meter is
ready to perform the test and is waiting for the
sample to be applied.
Simultaneously, a 5-minute countdown
begins. You must apply the sample within this
time, otherwise you will receive an error
message (after the sample detection has
timed out).
In case of errors, see “Troubleshooting” on
page 179.
QC Test - Sel. QC Level
Level 2
Level 1
proBNP PB1234
PBC014
JONES, TOM
Par:
Op:
Code:
QC:
04/19/2016
QC Test - Warming Up
Level: 1
proBNP, PB1234
PBC014
JONES, TOM
Par:
Op:
Code:
QC:
04/19/2016
Please Wait
W
arming Up
QC Test - App. QC
Apply QC
Min
150 µL
Level: 1
proBNP
PB1234
PBC014
JONES, TOM
Par:
Op:
Code:
QC:
4:10
04/19/2016

Table of Contents

Other manuals for cobas h 232

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the cobas h 232 and is the answer not in the manual?

cobas h 232 Specifications

General IconGeneral
Device TypePoint-of-care analyzer
Sample TypeWhole blood, plasma
Measured ParametersTroponin T, Myoglobin, D-Dimer, NT-proBNP
ConnectivityUSB
Intended Usecardiac markers
Data ManagementLIS connectivity
ThroughputSingle test
Sample Volume150 µL
Operating Temperature15-30 °C
Operating Humidity10-85% (non-condensing)
DisplayColor touchscreen
Measurement PrincipleElectrochemiluminescence immunoassay (ECLIA)