Control Testing and Quality Control
150
The result of this quality control test is dis-
played and is automatically saved to memory.
The target value and the range in which the
results for this control material should be
located is displayed below the current result.
If a quality control test fails, an up arrow (too
high) or down arrow (too low) is displayed.
There are three options for displaying a QC
test result:
Pass/Fail
Value
Value & Pass/Fail
For details refer to “QC result format” on
page 105 and “Settings summary” on
page 47.
QC Test - Result
329
04/19/2016 10:17
pg/mL
(350 / 300-400 pg/mL)
Level:
proBNP PB1234
PBC014
JONES, TOM
Par:
Op:
Code:
QC: 1
Pass:
04/19/2016
QC Test - Result
04/19/2016 10:17
(350 / 300-400 pg/mL)
Level:
proBNP PB1234
PBC014
JONES, TOM
Par:
Op:
Code:
QC: 1
04/19/2016
Fail:
490
pg/mL

Table of Contents

Other manuals for cobas h 232

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the cobas h 232 and is the answer not in the manual?

cobas h 232 Specifications

General IconGeneral
Device TypePoint-of-care analyzer
Sample TypeWhole blood, plasma
Measured ParametersTroponin T, Myoglobin, D-Dimer, NT-proBNP
ConnectivityUSB
Intended Usecardiac markers
Data ManagementLIS connectivity
ThroughputSingle test
Sample Volume150 µL
Operating Temperature15-30 °C
Operating Humidity10-85% (non-condensing)
DisplayColor touchscreen
Measurement PrincipleElectrochemiluminescence immunoassay (ECLIA)