Step Action
10 Close the lid of the instrument.
Proceed to the next section to start the formulation. It is recommended to start the
fo
rmulation as soon as possible after loading the syringes.
CAUTION
Do
not wait longer than 3 minutes between inserting the syringes
and starting the formulation. Leaving the heating block on with the
syringes inserted can cause ignitable vapor to form inside the
instrument.
6 Operation
6.
3 Run a procedure
6.3.6 Load syringes and collection tubes
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA 100