Step Action
7 Optional. Insert the L syringe into the left channel of the rotating block.
8 Secure the syringes to the cartridge by doing one of the following:
a.
Syringes with Luer fittings: Twist the syringe to engage the Luer connec-
tion.
NOTICE
Do
not tighten the Luer connection too much. Tight-
ening the connection too much can damage the
syringe and the cartridge.
b. Syringes without Luer fittings: Push the syringe until you meet resist-
ance.
NOTICE
Do
not insert the syringe too far. Inserting the syringe
too far can damage the syringe and the cartridge.
9 Install the sample collection tube in the right clip of the sample arm and the
wa
ste collection tube in the left clip of the sample arm.
Note:
Proper placement of the sample and collection tubes is critical to the avoid
unnecessary spillage.
If using the Ignite sample arms, or the 15 mL or 50 mL sample arms on
Ignite+: Make sure that the tubes sit as low as possible in the clips. The
threads of the tubes should sit directly above the plastic clips.
If using the 100 mL sample arm on Ignite+: Make sure the lip of the waste
t
ube is pressed upwards to sit as high as possible against the middle ridge
of the tube holder. Make sure the lip of the 100 mL sample tube is pressed
downwards to sit as low as possible on the top edge of the tube holder.
6 Operation
6.
3 Run a procedure
6.3.6 Load syringes and collection tubes
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA 99