Step Action
1 Rotate the rotating block upwards so that the underside of the cartridge is
visible.
Result:
2 Insert the empty syringes selected for the run into the heating inserts in the
R
and C channels of the rotating block.
Tip:
While waiting for the syringes to be preheated, change the sample arm if
necessary, and load the collection tubes.
3 Remove the syringes from the rotating block.
4 Load the preheated reagents into the syringes.
5 Remove any air bubbles at the tip of the syringes, making sure the fluid
me
niscus at the tip of the syringe is as flat as possible to prevent pre-mixing
of the fluids when loading the syringes.
Tip:
To prevent excess fluid from collecting in the Luer fitting, hold a lint-free labo-
ratory tissue near the tip of the syringe when expelling air bubbles.
6 Insert the R and C syringes into the heating inserts in the right and center
ch
annels, respectively.
6 Operation
6.
3 Run a procedure
6.3.6 Load syringes and collection tubes
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA 98