Dilution ratio
Th
e dilution ratio is the ratio between the volume of the dilution solution (L channel)
and the combined volume of the reagents from the C and R channels.
The value of the parameter Dilution ratio can be up to 10:1 for both instruments.
The limitations on the flow rate for each channel can reduce the maximum dilution
ratio for a specific recipe. The table below gives the maximum flow rate for the dilution
channel for each instrument. The dilution flow rate is calculated automatically by the
instrument based on the other parameter settings.
Instrument Dilution flow rate
Ignite
0.1 to 48 mL/min
Ignite+
0.1 to 435 mL/min
5 Preparation
5.
1 Parameters
5.1.2 Mixing parameters
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA 61