• Precision calibrated syringe detection is designed for optimal fluid injection into an
e
mpty fluid path.
• Start waste volumes are sufficient to wet and flush the channels prior to formula-
tion.
Follow the steps below to prime the system. If not using priming, proceed to Section
6.3.6 Load syringes and collection tubes, on page 94.
Step Action
1 Insert the waste collection tube into the left clip of the sample arm.
2 Fill syringes with the priming solvents. At least 6 mm (¼ inch) of fluid must be
ad
ded to each syringe.
3 Open the lid of the instrument.
4 Rotate the rotating block upwards so that the underside of the cartridge is
visible.
Result:
6 Operation
6.
3 Run a procedure
6.3.5 Prepare the cartridge
NanoAssemblr Ignite and Ignite+ Operating Instructions NIN1134 AA 91