DISPLAY MESSAGES AND FAULT FINDING
4-18 25257 Issue 3 November 2013
CAN-Bus Management Messages
No. Error Text Description
12005 Unknown SB0 during
news interpretation in
zyk task
Internal communication
Error list: System error message
Message category: Alarm
Remedial action: Restart control
12006 Unknown SB1 at
SB0_INIT
Internal communication
Error list: System error message
Message category: Alarm
Remedial action: Restart control
12007 Unknown SB1 at
SB0_EXCEPTION
Internal communication
Error list: System error message
Message category: Alarm
Remedial action: Restart control
12012 CAN-Drv: receive queue CAN-Driver, receive buffer overflow
Error list: Application message
Message category: Alarm
Remedial action: CAN-Bus is out of
order, Check CAN lines and load
resistance
12021 CAN-Drv: busof CAN-Driver, busof
Error list: Application message
Message category: Alarm
Remedial action: CAN-Bus is out of
order, Check CAN lines and load
resistance
12022 CAN-Drv: error CAN-Driver, error counter too high
Error list: Application message
Message category: Alarm
Remedial action: CAN-Bus is out of
order, Check CAN lines and load
resistance
12023 CAN-Drv receive error CAN-Driver, receive counter too high
Error list: Application message
Message category: Alarm
Remedial action: CAN-Bus is out of
order, Check CAN lines and load