Copyright 2008 Baker Hughes Company.
English–PACE5000/6000 Instruction Manual | iii
General Specification
Abbreviations
The following abbreviations are used in this manual. Abbreviations are the same in the singular
and plural.
Item Description
Display LCD: Color display with touch-screen.
Operating Temperature 10°C to 50°C (50°F to 122°F)
Storage Temperature -20°C to 70°C (-4°F to 158°F)
Ingress Protection IP20 (EN 60529)
Operating Humidity 5% to 95%
RH (non-condensing)
Vibration MIL-PRF-28800 Type 2 class 5 style E/F
Operating Altitude Maximum 2000 metres (6560 ft)
EMC EN 61326
Electrical Safety EN 61010-1, UL 61010-1, CSA 22.2, No. 61010-1 and IEC 61010-1
Power Supply PACE 5000: Input range: 100 - 240 V ac, (50/60 Hz), 2 A, installation
category
II, Fuse T2AH250V
PACE 6000: Input range: 100 - 120/200 - 240 V ac (50/60 Hz), 5A, installation
category
II, Fuse T5AH250V
Pressure Safety Pressure Equipment Directive - Class: Sound Engineering Practice (SEP) for
group
2 fluids.
Pollution Degree 2
Operating Environment Indoor use only. Do not use in potentially explosive environments.
Abbreviation Description
a Absolute
ac Alternating Current
dc Direct current
DPI Digital Pressure Instrument
etc. And so on
e.g. For example
ft Foot
g Gauge
GPIB General purpose interface bus
H
2
O Water
Hg Mercury
Hz Hertz
IDOS Intelligent Digital Output Sensor (Druck product)
i.e. That is