English–PACE5000/6000 Instruction Manual | 81
Communications - Instrument Emulation
The instrument should be returned to the manufacturer or calibration facility, see Section 6.22,
“Return Goods/Material Procedure,” on page
81.
To find the date of the last calibration, press Measure setup > Status > Calibration history.
6.20 Communications - Instrument Emulation
Refer to K0469 - PACE Communications Manual - Instrument Emulation.
6.21 Specification
Refer to PACE Module Pressure Controller data sheet.
Note: Data sheet 920-561 is contained in the CD-ROM shipped with the product.
6.22 Return Goods/Material Procedure
If the unit requires calibration or is unserviceable, return it to the nearest Druck Service Centre
listed at:
Druck.com
Contact the Service Department to obtain a Return Goods/Material Authorization (RGA or RMA).
Provide the following information for a RGA or RMA:
Product (e.g. PACE5000 or PACE6000).
Serial number.
Details of defect/work to be undertaken.
Include any error code(s) if available.
Operating conditions.
6.22.1 Safety Precautions
You must inform Druck if the product has been in contact with any hazardous or toxic substance.
The relevant COSHH or in the USA, MSDS, references and precautions to be taken when
handling.
6.23 Packaging Procedure
1. The instrument should be at zero/ambient pressure.
2. Switch off and isolate the electrical power supply to the instrument.
3. Shut off the pneumatic pressure and vacuum supplies to the instrument.
4. Remove the instrument from the equipment rack to access the rear panel.
5. Disconnect the power supply cable and the pneumatic supply hose assemblies.
6. Stow the power supply cable in the packaging below.
7. Remove any pressure adaptors, diffusers and restrictors.
Screen calibration -
Time & Date Sets instrument clock and calendar.
Change PIN Changes the Calibration PIN. Enter the existing PIN, the new PIN and
confirmation of the new PIN. Should this new PIN become lost it can only be
reset by returning the instrument to a Druck Service Center.
INFORMATION Service by unauthorized sources will affect the warranty and
may not guarantee further performance.
Item Description

Table of Contents

Other manuals for Druck PACE5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Druck PACE5000 and is the answer not in the manual?

Druck PACE5000 Specifications

General IconGeneral
BrandDruck
ModelPACE5000
CategoryController
LanguageEnglish