EasyManuals Logo

Druck PACE5000 User Manual

Druck PACE5000
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #77 background imageLoading...
Page #77 background image
Copyright 2008 Baker Hughes Company.
English–PACE5000/6000 Instruction Manual | 67
Barometric Reference Option
6.14 Barometric Reference Option
The barometric reference option measures the barometric pressure at the reference port.
When installed, this option allows absolute or gauge pressure range selection. To obtain absolute
pressure the instrument uses a summation of gauge pressure and barometric pressure
(measured by the barometric sensor).
11 CONTROL Controller ON
12 WAIT IN LIMITS Wait for In-Limits condition
13 BEEP ON Beep on, approximately 1 second, Beep off
14 MEASURE Switch to measure (controller off)
15 DWELL 30 Wait, 30 seconds
16 SET POINT 800 Set-point, 800
mbar
17 CONTROL Controller on
18 WAIT IN LIMITS Wait for In-limits condition
19 BEEP ON Beep on, approximately 1 second, Beep off
20 MEASURE Switch to measure, controller off
21 COUNT 1 Increment loop counter
22 VENT Vent
23 GOTO 9 Loop back to program line 9
24 BREAK
25 SET POINT 0
26 WAIT IN LIMITS
27 MEASURE
28 END Program end
Table 6-8: Example of Programming a Loop
Step Command Argument Action

Table of Contents

Other manuals for Druck PACE5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Druck PACE5000 and is the answer not in the manual?

Druck PACE5000 Specifications

General IconGeneral
BrandDruck
ModelPACE5000
CategoryController
LanguageEnglish