64 | PACE5000/6000 Instruction Manual–English
Chapter 6. Reference
GOTO Used to setup a loop. Enter program line number to go to.
IF If statement allows for a decision. For instance, if count 1 = 5 then GOTO xx
IN LIMITS In-Limits band setting (% full-scale).
IN LIMITS TIMER Waits, within In-limits for this time period, before setting a valid In-limits
condition and continuing the Test program sequence.
I/P LOGIC Specifies change of state for external contacts as a halt condition.
ISOLATION VALVE Control of output isolation valve: Isolated (closed) or Non-isolated (open).
MEASURE Selects Measure mode.
PAUSE Causes test program to pause and wait for user input (Resume).
RANGE Specifies instrument range. Test program can only run within one control
module.
RATE MAX Sets controller rate to maximum.
RATE VALUE Specifies controller linear rate entered., per minute or second
RESOLUTION Sets display resolution.
SET COUNT 1 Sets the counter 1 internal value.
SET COUNT 2 Sets the counter 2 internal value.
SETPOINT Allows set-point to be entered.
SETTLING FAST Used to specify controller fastest response, may overshoot.
SETTLING NO
OVERSHOOT
Used to specify controller response without overshoot.
STOP COUNT Used to stop loop program after a number of loops. (Legacy support only,
not recommended for use if count 1 and count 2 are listed).
TEXT Sets screen message.
UNITS Selects required display units.
VENT Instructs the module to vent.
VFC OFF Sets VFC allocation test program off. (Legacy support only, not
recommended for use if select VFC relay command is present).
VFC ON Sets VFC allocation test program on. (Legacy support only, not
recommended for use if select VFC relay command is present).
SELECT VFC RELAY 1 Allows control of VFC 1: R1, R2, R3 (When VFC is set to Test program under
VFC setup).
SELECT VFC RELAY 2 Allows control of VFC 2: R1, R2, R3 (When VFC is set to Test program under
VFC setup).
WAIT IN-LIMITS Waits until pressure is within the limits.
ZERO Output zeroed
Table 6-6: Test Program Commands
Command Description