Burst Pressure Testing Option
4. End slew rate, a default value of 1% full-scale can be changed to another value suitable for
the device under test. The slower the slew rate the more accurate the burst point detection
pressure.
User selected set-point near
the expected burst point.
Start slew rate default value
10% full-scale/second.
User selected set-point of the
expected burst point.
End slew rate default value 1%
full-scale/second.