Patient Monitor User Manual Monitoring ECG/ RESP
- 109 -
ST ANALYSE Menu
Figure 13-10 ST Analyse menu
ST Analysis Alarm Setting
ST ANALYSE: the switch for ST analysis. Set it to ON to activate the ST analysis or OFF
to disable the ST analysis.
ST ALM: pick ON to enable prompt message and data record during the ST analysis alarm;
pick OFF to disable the alarm function, and there will be a beside ST. ST alarm is
activated when the result exceeds set ST HI value or falls below ST LO value.
WARNING
In order to avoid endangering the patient’s life, the user should use this function
cautiously.
ALM LEV: used to set up the ST alarm level. There are three selections: HIGH, MED and
LOW.
ALM REC: pick ON to enable report printing upon ST analysis alarm.
ALM LIMIT SETUP: used to set up the upper limit and lower limit of ST alarm. The ALM
HI can be set to 0.2 mV to 2.0 mV, and the ALM LO can be set to –2.0 mV to 0. 2 mV. The
setup ALM HI should be higher than the ALM LO.
DEF POINT: pick this item to access the DEF POINT window, in which the position of
ISO and ST point can be set up.
ISO Base point. Default is 80 ms.
ST Measurement point. Default is 108 ms.