Patient Monitor User Manual Monitoring ECG/ RESP
- 110 -
Figure 13-11 DEF POINT Window
The operator can adjust the position of both ISO and ST measurement points.
Set the reference point of ST measurement point to be peak point of R-wave.
Figure 13-12 DEF POINT
The ST measurement for each beat complex is the vertical difference between the two
measurement points.
NOTE:
1 The ST measurement point should be adjusted if the patient's HR or ECG
morphology changes significantly.
2 The ST algorithm has been tested for accuracy of the ST segment data. The
significance of the ST segment changes need to be determined by a clinician.
Adjusting ISO, ST:
These two points can be adjusted by turning the knob.
When adjusting ST measurement point, the system will show the ST Measurement Point Window.
ST value