Patient Monitor User Manual Intended Use and Safety Guidance
- 4 -
WARNING
17 Devices connecting with monitor should be equipotential.
18 When the monitor and electrosurgical device are used together, the user (physician
or nurse) should guarantee the safety of patient.
19 This equipment is not intended for family usage.
20 Disposable devices are intended for single use only. They should not be reused as
performance could degrade or contamination could occur.
21 Without use of data store function, all data measured (including trend data, review
data, alarm events and so on) are cleared either when the monitor is turned off or
when the monitor is powered down in the process of monitoring.
22 Additional multiple socket-outlet or extension cord cant be connected to the system.
23 Only items that have been specified as part of the system or specified as being
compatible with the system can be connected to the system.
24 SHOCK HAZARD - Don't connect electrical equipment, which has been supplied as a
part of the system, directly to the wall outlet when the non-medical equipment is
intended to be supplied by a multiple portable socket-outlet with an isolation
transformer.
25 SHOCK HAZARD - Don't connect electrical equipment, which has not been supplied
as a part of the system, to the multiple portable socket-outlet supplying the system.
26 Do not touch accessible parts of medical or non-medical electrical equipment in the
patient environment and the patient simultaneously, such as USB connector, VGA
connector or other signal input/output connectors.
27 Connecting any accessory (such as external printer) or other device (such as the
computer) to this monitor makes a medical system. In that case, additional safety
measures should be taken during installation of the system, and the system shall
provide:
a) Within the patient environment, a level of safety comparable to that provided by
medical electrical equipment complying with IEC/EN 60601-1, and
b) Outside the patient environment, the level of safety appropriate for non-medical
electrical equipment complying with other IEC or ISO safety standards.
28 All the accessories connected to system must be installed outside the patient vicinity,
if they do not meet the requirement of IEC/EN 60601-1.
29 Using accessories other than those specified may result in increased electromagnetic
emission or decreased electromagnetic immunity of the monitoring equipment.