Patient Monitor User Manual Monitoring IBP (Optional)
- 139 -
Pick the IBP SETUP item to call up the IBP SETUP menu shown as follows:
Figure 17-3 IBP SETUP Menu
The items to be set up in the menu include:
ALM: select ON to enable alarm prompt during IBP alarm. Select OFF to disable audio
alarm and prompt the symbol beside IBP numeric.
WARNING
In order to avoid endangering the patient’s life, the user should use this function
cautiously.
ALM LEV: set the alarm level. Two levels are available: HIGH and MED.
ALM REC: select ON to enable recording during the IBP alarm or to OFF to disable the
alarm recording function.
AMP ADJUST: set to adjust waveform amplitude. Two selections are available: MANUAL,
AUTO. Set it to AUTO, the pressure names of IBP become P1 and P2, and the IBP scale is
adjusted by system automatically. Set it to MANUAL, the pressure names of IBP can choose
one of ART, PA, CVP, RAP, LAP, ICP, P1, P2 and the IBP scale is adjusted by the user
via SCALE ADJUST item.
SWEEP: set to select the scanning speed of the IBP wave. Four selections are available:
6.25mm/s, 12.5 mm/s, 25 mm/s or 50.0mm/s.
UNIT: set to select the pressure unit (mmHg, kPa, or cmH2O).
FILTER: set this item to 12.5Hz or 40.0 Hz.
ALM LIMIT SETUP: used to access the sub-menu of IBP ALM LIMIT SETUP, in which
the user may set up the upper and lower alarm limits of systolic pressure, diastolic pressure
and mean pressure respectively for channel 1 and channel 2.