Patient Monitor User Manual Specifications
- 177 -
Measurement
method
Non-dispersive infrared gas analysis (NDIR)
Unit
mmHg, %, kPa
Measuring
Range
CO
2
0 mmHg to 150 mmHg (0% to 20%)
AwRR
2 rpm to 150 rpm
Resolution
EtCO
2
1 mmHg
FiCO
2
1 mmHg
AwRR
1 rpm
Accuracy
EtCO
2
2
mmHg, 0
mmHg to
40 mmHg
Respiratory
rate: 60
rpm
Typical
conditions:
Ambient
temperature:
25 C 3
C
Barometric
pressure:
760 mmHg
10 mmHg
Balance
gas: N
2
Sample gas
flowrate:
100 ml/min
5% of
reading,
41 mmHg
to 70
mmHg
8% of
reading,
71 mmHg
to 100
mmHg
10% of
reading,
101
mmHg to
150
mmHg
12% of
reading or
4
mmHg,
whichever
is greater
Respiratory
rate: > 60
rpm
AwRR
1 rpm
Drift of
measurement
accuracy
Meets the measurement accuracy above.
Pump rate
70 ml/min or 100 ml/min, (Accuracy: 15 ml/min). Default: 100 ml/min

Table of Contents

Related product manuals