Patient Monitor User Manual Specifications
- 178 -
Warm-up
time
Display waveform within 20 s; Reach the design accuracy within 2 minutes.
Rise time
< 400 ms (with 2 m gas sampling tube, sample gas flowrate: 100 ml/min)
< 500 ms (with 2 m gas sampling tube, sample gas flowrate: 70 ml/min)
Response
time
< 4 s (with 2 m gas sampling tube, sample gas flowrate: 100 ml/min)
< 4 s (with 2 m gas sampling tube, sample gas flowrate: 70 ml/min)
Work mode
Standby, measure. Default: Standby
O
2
compensation
Range: 0% to 100%
Resolution: 1%
Default: 16%
N
2
O
compensation
Range: 0% to 100%
Resolution: 1%
Default: 0%
AG
compensation
Range: 0% to 20%
Resolution: 0.1%
Default: 0%
Humidity
compensation
method
ATPD (default), BTPS
Barometric
pressure
compensation
Automatic (Barometric pressure change will not bring additional deviation to
measuring accuracy).
Calibration
Support
Alarm
EtCO
2
, FiCO
2
, AwRR
Apnea
alarm delay
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
Data Sample
Rate
100 Hz
EtCO
2
Change
1
AwRR 80rpm, meet the accuracy
mentioned above;
AwRR > 80 rpm, EtCO
2
descends
8%;
AwRR > 120 rpm, EtCO
2
descends
10%
with 2 m gas sampling tube, sample gas
flowrate: 100 ml/min)