Patient Monitor User Manual Specifications
- 179 -
AwRR ≤ 60rpm, meet the accuracy
mentioned above;
AwRR > 60 rpm, EtCO
2
descends
8%;
AwRR > 90 rpm, EtCO
2
descends
10%;
AwRR > 120rpm, EtCO
2
descends
15%;
with 2 m gas sampling tube, sample gas
flowrate: 70 ml/min)
Note 1: Use a test device equivalent to EN ISO 80601-2-55 fig 201.101 to measure at 1:2 I/E
ratio. Respiration rate accuracy is determined by frequency of device, and end-tidal reading
change refers to the nominal value.
Respironics Module
Adult, pediatric and neonatal patients
Infra-red Absorption Technique
2 rpm to 150 rpm
(Sidestream)
0 rpm to 150 rpm
(Mainstream)
2 mmHg, 0 mmHg to 40 mmHg
5% of reading, 41 mmHg to 70 mmHg
8% of reading, 71 mmHg to 100 mmHg
10% of reading, 101 mmHg to 150 mmHg
12% of reading, RR is over 80 rpm (Sidestream)
There will be no degradation in performance due to
Respiration Rate. (mainstream)