Patient Monitor User Manual Specifications
- 180 -
AwRR
1 rpm
Sample Gas Flow Rate
(Sidestream)
50 ml /min 10 ml /min
O
2
Compensation
Range
0% to 100%
Resolution
1%
Default
16%
Anesthetic Gas Compensation
Range
0% to 20%
Resolution
0.1%
Default
0.0%
Balance Gas Compensation
Options: N
2
O, helium, room air
Barometric pressure
compensation
User setup
Work Mode
Standby, measure. Default: Standby
Stability
Short Term Drift
< 0.8 mmHg over 4 hours
Long Term Drift
Accuracy specification will be maintained over 120
hours period
Initialization time
It displays the value within 15 s and meets the
requirement for measurement accuracy within 2 min.
(Mainstream)
It displays the value within 20 s and meets the
requirement for measurement accuracy within 2 min.
(Sidestream)
Response time
60 ms (Mainstream)
Calibration
Not required.
Alarm
EtCO
2
, FiCO
2
and AwRR alarm
Apnea Alarm Delay (Unit: s)
10, 15, 20, 25, 30, 35, 40; 20 s by default