Patient Monitor User Manual Alarm Information
- 68 -
RESP adjustable range of alarm limits are listed as follows (unit rpm):
Patient Type
ALM HI
ALM LO
ADU
120
6
PED
150
6
NEO
150
6
SpO
2
adjustable range of alarm limits are listed as follows (unit %):
ALM HI
ALM LO
SpO
2
100
20
PR
adjustable range of alarm limits are listed as follows (unit bpm):
ALM HI
ALM LO
PR (SpO
2
)
300
30
PR (NIBP)
240
40
PR (IBP)
300
30
NIBP adjustable range of alarm limits are listed as follows (unit: mmHg, kPa; 1 mmHg=0.133
kPa):
Patient Type
ALM HI
ALM LO
ADU
SYS
DIA
MAP
270
215
235
40
10
20
PED
SYS
DIA
MAP
230
180
195
40
10
20
NEO
SYS
DIA
MAP
135
100
110
40
10
20
TEMP adjustable range of alarm limits are listed as follows:
ALM HI
ALM LO
T1
50 C (122 )
0 C (32 )
T2
50 C (122 )
0 C (32 )
TD
high limit: 0.1 C (32.18 )~50 C (122 )