Patient Monitor User Manual Alarm Information
- 69 -
IBP adjustable range of alarm limits are listed as follows (unit mmHg):
ALM HI
ALM LO
Art
300
0
RAP
40
-10
LAP
40
-10
ICP
40
-10
CVP
40
-10
PA
120
-6
P1
300
-50
P2
300
-50
CO
2
adjustable range of alarm limits are listed as follows:
ALM HI
ALM LO
EtCO
2
150 mmHg
0 mmHg
FiCO
2
50 mmHg
0 mmHg
AwRR
150 rpm
2 rpm