Patient Monitor User Manual Maintenance/ Cleaning
- 90 -
Replacing the Air Bladder:
After cleaning, replace the air bladder into the cuff following the steps below:
1. Roll the bladder lengthwise and insert it into the cuff from the large opening at one end of
the cuff.
2. Thread the hose from within the cuff and out through the small hole at the top of the cuff.
3. Adjust the bladder until it is in position.
Cleaning the SpO
2
Sensor
1. Wipe the surfaces of the sensor and cable using a soft cloth dampened with the cleaning
solution until no visible contaminants remain.
2. Wipe the patient contact area of the sensor with the cotton swab dampened with the cleaning
solution. until no visible contaminants remain
3. Wipe off the cleaning solution with a fresh cloth or towel, dampened with tap water after
cleaning until no visible cleaning agent remains.
4. Wipe off with a dry cloth to remove residual moisture.
5. Leave the sensor to air dry.
Cleaning the IBP Cable
1. Wipe the cables with a soft cloth dampened with the cleaning solution until no visible
contaminants remain.
2. Wipe off the cleaning solution with a fresh cloth or towel, dampened with tap water after
cleaning until no visible cleaning agent remains.
3. Wipe off with a dry cloth to remove residual moisture.
4. Leave the cables to air dry.
Cleaning the TEMP Sensor
1. Wipe the patient contact area with a soft cloth dampened with the cleaning solution until no
visible contaminants remain.
2. Wipe off the cleaning solution with a fresh cloth or towel, dampened with tap water after
cleaning until no visible cleaning agent remains.
3. Wipe off with a dry cloth to remove residual moisture.
4. Leave the sensor to air dry.
12.3.2.3 Cleaning the Dust-proof Filter Assembly
The dust-proof filter system consists of two parts, one is the filter base near the battery door of
device (as left picture shown), and another is the dust proof filter assembly (as right picture
shown) between the front housing and rear housing. Through filtering the dust in air and the