To check what Standby Speed is set, send a query as follow:
Command ? S 8 5 7 cr
The reply is as follows:
Reply = S 8 5 7 sp d d d cr
5.1.7 Normal speed seng
The normal speed is a user-congurable seng and can be set to the value between
50% and 100% full rotaonal speed. When the pump is shipped, it is congured with a
normal speed of 80% full rotaonal speed.
To change the normal speed seng, send the command that follows (characters 'd'
shows the value as a percentage of full rotaonal speed):
Command ! S 8 5 6 sp d d d cr
The reply is as follows:
Reply * S 8 5 6 sp r cr
The normal speed is now stored in the memory of the pump.
To check what normal speed is set, send a query as follows:
Command ? S 8 5 6 cr
The reply is as follows:
Reply = S 8 5 6 sp d d d cr
5.1.8 Timer seng and opons
The meout period is a user-congurable opon and can be set to the value from
1 to 30 minutes.
When the pump is shipped, it is congured with a default meout period of 8 minutes.
Refer to Timer on page 12 for a full descripon of mer funconality.
To change the mer seng, send the command that follows (where the characters 'd'
represent the meout period in minutes):
Command ! S 8 5 4 sp d d cr
The reply will be as follows:
Reply * S 8 5 4 sp r cr
The Timer seng is now stored in the memory of the pump.
To check what meout period is set, send a query as follows:
Command ? S 8 5 4 cr
The reply will be as follows:
Reply = S 8 5 4 sp d d cr
Page 62
B8G000880_D - Conguraon