Error ash
posion
Error ash
code
Comments Acons
4 sssLss
Overspeed or overcurrent trip is
acvated, or other hardware fault
Restart the system. If the error
code is not cleared, contact us or
the supplier.
5 ssssLs
Pump internal measurement
system is disconnected or
damaged
Restart the system. If the error
code is not cleared, contact us or
the supplier.
6 sssssL
Serial enable becomes inacve
aer a Serial Start command
Acvate the Serial Enable again
and send a Serial Stop command
to clear the error code.
Note:
The alarm LED error ash sequence is capable of signaling mulple fail condions. For
example, error ash code sLssLs signies error 2 (controller failed internal conguraon
and calibraon operaon) and error 5 (pump internal measurement system disconnected
or damaged).
8.4 Decoding system status words
When you use the serial communicaons link, addional informaon that can be useful
for the fault nding can be accessed. When sending a query to monitor measured motor
speed, the pump also returns a System Status Word.
The send command that follows:
Command ? v 8 5 2 cr
The reply is as follows (where, rst returned number refers to motor rotaonal speed in
revoluons per second (Hz) and second part is the system status word):
Reply = v 8 5 2 sp d d d d ; h h h h h h h h cr
The System Status Word returned is made up of 8 hexadecimal digits. To decode this
word, convert each digit into a 4‑digit binary number. Example:
2 2 8 3 0 0 2 2
ê ê ê ê ê ê ê ê
0 0 1 0 0 0 1 0 1 0 0 0 0 0 1 1 0 0 0 0 0 0 0 0 0 0 1 0 0 0 1 0
Refer to Table: Hexadecimal conversion table on page 91 for more informaon.
Table 32
Hexadecimal conversion table
Hexadecimal Binary Decimal
0 0000 0
1 0001 1
2 0010 2
3 0011 3
4 0100 4
5 0101 5
Page 91
B8G000880_D - Fault nding