The reply will be as follows:
Reply = S 8 7 1 sp d cr
5.1.10 Electronic braking opons
Refer to Electronic braking on page 14 for a full descripon of the electronic braking
feature. To enable Electronic Braking, send the serial command that follows:
Command ! S 8 7 2 sp 1 cr
The reply will be as follows:
Reply * S 8 7 2 sp r cr
The state of the Electronic Braking opon is stored in the memory of the pump.
To disable the Electronic Braking again, send the serial command that follows:
Command ! S 8 7 2 sp 0 cr
The reply will be as follows:
Reply * S 8 7 2 sp r cr
To check if electronic braking is enabled or disabled, send the query that follows:
Command ? S 8 7 2 cr
The reply will be as follows (where d=0 means disabled and d=1 means enabled):
Reply = S 8 7 2 sp d cr
5.1.11 Factory sengs
The pump can be congured again to its original factory sengs with one serial
command.
To reset the motor controller to factory sengs, send the command that follows:
Command ! S 8 6 7 sp 1 cr
The reply will be as follows:
Reply * S 8 6 7 sp r cr
The factory sengs are restored in the memory of the pump. If the valve type seng
was previously congured to a non-default seng, a Power cycle can be necessary.
5.1.12 Assigning a mul-drop address
Each individual pump must be programmed with its own mul-drop address through a
point‑to‑point connecon before introducon into a mul-drop network.
When the pump is shipped, mul-drop mode is disabled by default.
To assign a mul‑drop address, send the command that follows (where the 'd' characters
represent the address):
Page 64
B8G000880_D - Conguraon