WARNING: AUTOMATIC RESTART
Risk of injury. When the power is restored aer a power cut, the pump will restart
automacally. Keep the pump connected to the vacuum system to prevent risk of
injury.
CAUTION: PUMP MOVEMENT
Risk of damage to equipment. Do not move the pump when it is in operaon. The
gyroscopic forces generated by this movement can cause more use of the back up
bearing and can result in catastrophic failure of the pump.
In serial control, the pump will accelerate to full operang speed when started.
To start the pump, send the command (over the serial communicaons link) that follow:
Command ! c 8 5 2 sp 1 cr
The reply is as follows:
Reply * c 8 5 2 sp 1 cr
The green indicator LED illuminates when the pump reaches normal speed. This is 80%
of the full rotaonal speed by default, but a dierent value can be selected to suit the
applicaon.
6.4.3 Run the pump at standby speed
If the pump operates at less than or more than the standby speed it will accelerate or
decelerate unl standby speed is reached.
To operate the pump at standby speed, send the command (over the serial
communicaons link) that follows:
Command ! c 8 6 9 sp 1 cr
The reply is as follows:
Reply * c 8 6 9 sp r cr
To return the pump to full speed, send the command that follows:
Command ! c 8 6 9 sp 0 cr
The reply is as follows:
Reply * c 8 6 2 sp r cr
6.4.4 Stop the pump
When the stop command is received, the pump rotor will decelerate and stop.
To stop the pump, send the command (over the serial communicaons link) that follows:
Command ! c 8 5 2 sp 0 cr
The reply is as follows:
Page 74
B8G000880_D - Operaon